1041897, R
EV
. 00
P
AGE
42 - E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
To install the Pressure Regulator
1. Apply a layer of Teflon thread tape to the threads of the pressure regulator.
2. Align the pressure regulator with the pressure regulator port in the Sieve Canister Assembly.
Screw the pressure regulator into the Sieve Canister Assembly until strong resistance is felt and
the pressure regulator outlet port is oriented in the correct position.
3. Install the pressure tubing on the pressure regulator outlet port.
4. Connect the unit to a power source and turn on the unit. Apply snoop leak detector to the area
where the pressure regulator mounts to the Sieve Canister Assembly and where the O
2
tubing
connects to the outlet barb to check for leaks.
CAUTION
The tape should be installed in a counter clockwise
direction, with the Pressure Regulator threads facing
you.