1041896, R
EV
. 00
P
AGE
38 - M600/M605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
5. Lift the Valve/Solenoid Assembly up and away from the Sieve Canister Assembly.
F
IGURE
AL: R
EMOVAL
OF
THE
V
ALVE
/S
OLENOID
A
SSEMBLY
To install the Valve/Solenoid Assembly
1. Align the Valve/Solenoid Assembly with the holes in the top of the Sieve Canister Assembly.
2. Using the five screws and flat washers provided, secure the Valve/Solenoid Assembly to the Sieve
Canister Assembly. The screws should be torqued sufficiently in order to prevent leaks, but care
should be taken not to crush the shoulder of the valve.
3. Reconnect the two wire connectors to the Valve/Solenoid Assembly.
4. Connect the unit to a power source and turn on the device. Apply snoop leak detector to the area
where the Valve/Solenoid Assembly mounts to the Sieve Canister Assembly and check for leaks.