To install the Power Switch Wiring Harness
1. Either feed the Power Switch wiring harness through the original screw down cable ties, or if new
screw down cable ties are available, loosely install four cable ties grouping the Power Switch wir-
ing harness, the O
2
/solenoid/battery wiring harness, and the clear and yellow pressure tubing
together.
2. Install the connectors onto the proper terminals of the Power Switch. Ensure that they are
installed in their original position.
F
IGURE
T: P
OWER
S
WITCH
T
ERMINAL
L
OCATIONS
3. Install the female connector onto the (J2) location on the Main PCB.
4. Position the screw down cable ties at their mounting posts then secure each using the original
screws. Tighten the screw down cable ties.
5. Connect the male connector on the Power Switch wiring harness from the female connector on the
main power wiring harness.
Red
Red
White
Black
Yellow/
Brown
Yellow/
Blue