1041896, R
EV
. 00
P
AGE
46 - M600/M605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
B
ASE
C
ABINET
R
EPLACEMENT
To remove the Base Cabinet
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove Compressor Cover/Perforated Canopy. Refer to the Compressor Cover Replacement
Section.
4. Remove the Capacitor. Refer to the Motor Start/Run Capacitor Section.
5. Remove the Compressor/motor assembly. Refer to the Compressor Assembly Replacement Sec-
tion.
6. Remove the In-line Muffler. Refer to the In-line Muffler Replacement Section.
7. Remove the Hour Meter. Refer to the Hour Meter Replacement Section.
8. Remove the Sieve Canister Assembly. Refer to the Sieve Canister Assembly Replacement Sec-
tion.
9. Remove the Power Cord. Refer to the Power Cord Replacement Section.
10. Remove the Base Cabinet and bottom pan.
To install the Base Cabinet
1. All components have now been removed from the original Base Cabinet. To begin reassembly,
place the Base Cabinet on a protected work surface. Follow the disassembly process in reverse
order to reassemble the unit.
NOTE
If care is taken during the following procedures, the Valve/Solenoid Assembly and the pressure
regulator do not have to be removed. However, if these components are damaged during the
procedure, they must be replaced.