To remove the OPI Wiring Harness
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the OPI wiring harness connector from the J3 location on the Main PCB assembly by pull-
ing the connector directly up from the J3 receptacle.
4. Remove the OPI wiring harness connector from the J7 location on the OPI board by pulling the
connector directly up from the J7 receptacle.
F
IGURE
Q: OPI W
IRING
H
ARNESS
To install the OPI Wiring Harness
1. Align the connector on the OPI board wiring harness with the J7 receptacle on the OPI board.
Carefully press the connector onto the receptacle until completely seated.
2. Align the connector on the OPI wiring harness with the J3 receptacle on the Main PCB assembly.
Carefully press the connector onto the receptacle until completely seated.
WARNING
Make sure the AC power is disconnected.