To remove the Sieve Canister Assembly
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
4. Remove the Valve/Solenoid Assembly. Refer to the Valve/Solenoid Replacement Section.
5. Remove the pressure regulator. Refer to the Pressure Regulator Replacement Section.
6. Using the 5/16” Nut Driver (Supplied in RI p/n H647, Millennium Tool Kit), remove the four screws
securing the Sieve Canister Assembly to the Base Cabinet.
7. Using a crimping clamp tool or diagonals, cut the one-eared clamp securing the 12” pressure tub-
ing to the inlet port of the Sieve Canister.
8. Remove the 12” pressure tubing from the inlet port of the Sieve Canister.
F
IGURE
AN: L
OCATION
OF
THE
O
NE
-
EARED
C
LAMP
AND
M
OUNTING
S
CREWS
(3
OF
4 S
HOWN
)
9. Lift the Sieve Canister Assembly up and out of the Base Cabinet.
10. Remove the Blow Down Muffler from the Sieve Canister Assembly (refer to the Blow Down Muffler
Replacement procedures).
11. Remove the brass fitting from the top of the Sieve Canister Assembly.
To install the Sieve Canister Assembly
1. Install the supplied canister gasket on the bottom of the Sieve Canister Assembly.
2. Install the brass fitting in the top of the Sieve Canister Assembly.
3. Install the Blow Down Muffler onto the Sieve Canister Assembly. Refer to the Blow Down Muffler