E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
45
1041897, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
B
LOW
D
OWN
M
UFFLER
R
EPLACEMENT
To remove the Blow Down Muffler
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
4. Remove the Valve/Solenoid Assembly. Refer to the Valve/Solenoid Replacement Section.
5. Remove the Pressure Regulator. Refer to the Pressure Regulator Replacement Section.
6. Remove the Sieve Canister Assembly.
7. Note the orientation of the blown down muffler in relation to the Sieve Canister Assembly.
8. Using channel locks, turn the Blow Down Muffler counter-clockwise until it is free of the sieve
canister assembly.
F
IGURE
AM: R
EMOVING
THE
B
LOW
D
OWN
M
UFFLER
NOTE
If care is taken during the following procedures, the Valve/Solenoid Assembly and the pressure
regulator do not have to be removed. However, if these components are damaged during the