14. Place the new piston cup on the back of the new retainer plate. Using the new screws provided,
secure the retainer to the connecting rod assembly and torque screws to 30 in.-lbs.
F
IGURE
N: I
NSTALLING
THE
P
ISTON
C
UP
& R
ETAINER
15. Set the new sleeve in place with the O-ring away from the piston. Slide the sleeve down over the
piston cup using your fingers to guide the sleeve without damaging the piston cup.
F
IGURE
O: I
NSTALLING
THE
C
YLINDER
S
LEEVE