1041898, R
EV
. 01
P
AGE
2 - M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
O
VERLAY
R
EPLACEMENT
To remove the Overlay
1. Starting at one corner, carefully peel the Overlay off of the cabinet.
2. Use a small amount of isoprppyl alcohol applies to a cleaning cloth to remove any adhesive resi-
due from the cabinet.
To install the Overlay
1. Remove the protective backing from the Overlay.
2. Carefully align the Overlay with the applicable mounting area. Set the Overlay in place and, using
a circular motion, firmly rub the Overlay until it is securely attached to the cabinet.
F
IGURE
A: P
ROPER
I
NSTALLATION
OF
F
RONT
C
ABINET
O
VERLAY
NOTE
When using isopropyl alcohol to clean the mounting area for the control Overlay,
ensure that no liquid enters the Front Cabinet or comes into contact with the
LEDs. Use protective glasses.