E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
7
1041897, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
F
LOW
M
ETER
R
EPLACEMENT
To remove the Flow Meter
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Working from inside the Front Cabinet, cut the cable ties then remove the pressure tubing from
both fittings on the Flow Meter by pulling the tubing from each fitting.
F
IGURE
G: V
IEW
OF
F
LOW
M
ETER
F
ROM
I
NSIDE
THE
F
RONT
C
ABINET
4. While holding the Flow Meter in place, remove the two speed nuts from the threaded fittings on the
back of the Flow Meter.
5. Remove the Flow Meter from the Front Cabinet assembly by pulling it straight out from the Front
Cabinet.
NOTE
The pressure tubing connected to the top fitting on the Flow Meter is connected to the bacteria filter.
The pressure tubing connected to the bottom fitting is connected to the OPI sensor for Model 605, or
the 1/2” tubing in the wiring harness for Model 600.