This chapter illustrates the repair and replacement procedures for the components of the Millennium Oxygen
Concentrators. Once repair is complete refer to the testing section for proper testing procedures.
WARNINGS
•
To prevent electrical shock, disconnect the electrical supply before attempting to make any
device repairs to the Millennium Oxygen Concentrators.
•
Use extreme caution when handling the Compressor/motor assembly Capacitor. The Capacitor
can hold a substantial electrical charge until it is properly discharged.
•
Avoid handling the molecular sieve material. Respironics recommends the return of the Sieve
Canister Assembly to Respironics for any service that involves sieve material disposal.
•
Even with the unit disconnected from the electrical supply, the Capacitor may still hold an
electrical charge strong enough to cause serious bodily injury. DO NOT touch the Capacitor
terminals simultaneously until the Capacitor has been completely discharged. Discharge the
Capacitor by shorting the two posts with an insulated screw driver.
CAUTIONS
•
Electronic components used in this unit are subject to damage from static electricity. Repairs
made to this unit must be performed only in an antistatic, ESD-protected environment.
•
During all repair and replacement procedures, assure that any connections that are broken
during the process (fittings, tubing, and hoses) are reconnected securely.
•
When using leak detector, be careful that it does not come in contact with any electrical
components.
•
The fuse on the power control board must be replaced with a fuse of the same rating as the
original. Failure to do so will result in damage to the unit.