1041897, R
EV
. 00
P
AGE
30 - E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
To install the Compressor Assembly
1. Align the Compressor/motor assembly with the standoffs and helical springs. Set the Compressor
assembly in place.
2. If M10600/M10605 Enhanced Compressor was mounted with 4 rubber bumpers, reinstall the 4
bumpers and 4 screws securing the Compressor to the Base Cabinet.
3. Slide the supplied one-eared clamp onto the pressure tubing. Align the tubing with the Compres-
sor assembly outlet port. Slide the tubing onto the ports.
4. Slide the one-eared clamp into position on the pressure tubing. Using a crimping clamp tool or
diagonals, firmly “pinch” the one-eared clamp to secure the pressure tubing.
5. Confirm that the In-line Muffler is not touching the side wall of the Base Cabinet
6. Install the Capacitor wire connectors onto the terminals of the Capacitor.
7. Push together the brown wire connector from the Compressor to the white wire connector from the
Main PCB assembly. Press the connectors together until completely seated. Refer to the block
diagram in Figure Y.
8. Push together the blue wire connector from the thermal protection switch to the black wire connec-
tor from the Main PCB assembly. Press the connectors together until completely seated. Refer to
the block diagram in Figure Y.