M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
29
1041898, R
EV
. 01
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
M
AIN
P
OWER
W
IRING
H
ARNESS
R
EPLACEMENT
To remove the Main Power Wiring Harness
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
4. Disconnect the female connector on the main power wiring harness from the male connector on
the Power Switch harness.
F
IGURE
AC: L
OCATION
OF
THE
M
AIN
P
OWER
W
IRING
H
ARNESS
5. Note the position of the Power Cord wires in the main power harness female connector. Using an
amp terminal retractor, remove the Power Cord wire from the connector.
6. Note the routing of the main power wiring harness and the location of the cable ties securing the
harness to the Base Cabinet. Cut the cable ties.
7. Remove the two connectors from the Hour Meter.
8. Disconnect the female connector of the main power wiring harness from the male connector on the
Compressor wires.
WARNING
Make sure the AC power is disconnected.
NOTE
The wires connected to the Hour Meter are interchangeable.