1041895, R
EV
. 00
P
AGE
22 - H600/H605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
4. Looking at the back of the Power Switch from inside the Front Cabinet, note the position and col-
ors of the wires connected to the terminals on the Power Switch.
5. Using needle nose pliers, remove the connectors from the Power Switch terminals.
6. Remove the screws that secure the screw down cables ties to the Front Cabinet.
7. Either feed the Power Switch wiring harness through the screw down cable ties or, if new screw
down cable ties are available, cut the cable ties and remove the Power Switch wiring harness.
To install the Power Switch Wiring Harness
1. Either feed the Power Switch wiring harness through the original screw down cable ties, or if new
screw down cable ties are available, loosely install four cable ties grouping the Power Switch wir-
ing harness, the O
2
/solenoid/battery wiring harness, and the clear and yellow pressure tubing
together.
2. Install the connectors onto the proper terminals of the Power Switch. Ensure that they are
installed in their original position.
F
IGURE
Y: P
OWER
S
WITCH
T
ERMINAL
L
OCATIONS
CAUTION
Except for the two red wires that are interchangeable, the wires must be connected
to the Power Switch as shown. Failure to do so will result in damage to the unit.
NOTE
The two red wires from the O
2
/solenoid/battery wiring harness do not have to be
disconnected from the Power Switch.
Red
Red
White
Black
Yellow/
Brown
Yellow/
Blue