M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
27
1041898, R
EV
. 01
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
3. Connect the fan wires to the terminals on the side of the cooling fan.
P
ERFORATED
C
ANOPY
R
EPLACEMENT
To remove the Perforated Canopy
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the 9-volt alarm battery from its holder.
4. Press inward on the bottom of each side of the Perforated Canopy until the locating posts are clear
of the holes in the Compressor Cover.
5. Lift the Perforated Canopy up and away from the Compressor Cover.
To install the Perforated Canopy
1. While exerting slight inwards pressure on the sides of the Perforated Canopy, align the locating
posts with the holes in the Compressor Cover.
2. Release the pressure, then ensure that the locating posts have locked in place in the Compressor
Cover.
3. Install the 9-volt battery in its holder.
F
IGURE
AA: P
ERFORATED
C
ANOPY
NOTE
Ensure that the convoluted tubing covering the fan wires is still properly seated in the
wire holder built into the front of the Compressor.