1041898, R
EV
. 01
P
AGE
32 - M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
C
OMPRESSOR
A
SSEMBLY
R
EPLACEMENT
To remove the Compressor Assembly
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
F
IGURE
AF: C
OMPRESSOR
A
SSEMBLY
S
EATED
IN
B
ASE
C
ABINET
4. Disconnect the Compressor assembly harness from the main power cabinet wiring harness by
depressing the locking tabs and separating the connectors.
5. Disconnect the wires from the Capacitor terminals.
6. Using a crimping clamp tool or diagonals, cut the one-eared clamp securing the pressure tubing to
the outlet port of the Compressor assembly. Remove the pressure tubing from the Compressor
assembly outlet port.
7. Using a 1/8” hex key wrench or socket, remove the four shoulder screws and rubber bumpers
securing the Compressor assembly to the standoffs.
8. Lift the Compressor assembly up from the standoffs and helical springs.
WARNING
Ensure that the unit is disconnected from the AC power source before beginning this procedure.
NOTE
The wires connected to the Capacitor are interchangeable.
NOTE
If the optional Compressor mount kit is being used, remove the four
helical springs from the Compressor assembly standoffs.