1041897, R
EV
. 00
P
AGE
40 - E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
To install the Valve/Solenoid Assembly
1. Align the Valve/Solenoid Assembly with the holes in the top of the Sieve Canister Assembly.
2. Using the five screws and flat washers provided, secure the Valve/Solenoid Assembly to the Sieve
Canister Assembly. The screws should be torqued sufficiently in order to prevent leaks, but care
should be taken not to crush the shoulder of the valve.
3. Reconnect the wire connectors to terminals J8, J9, J11, and J12 on the Main PCB assembly.
Refer to the block diagram in Figure AG.
4. Connect the unit to a power source and turn on the device. Apply snoop leak detector to the area
where the Valve/Solenoid Assembly mounts to the Sieve Canister Assembly and check for leaks.