S
PECIFICATIONS
& F
EATURES
- P
AGE
3
1041848, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
SAFETY
Pressure Relief Valve
40 ± 3 psig (303 ± 20.7 kPa)
Compressor Compartment Thermal
Protection with Self Reset
149° F (65° C)
ALARM CONDITIONS
Visual Red Light Emitting Diode (LED) and audible sound alert:
•
Start-up test;
•
Power failure;
•
Main PCB failure; and
•
High and low pressure
Oxygen percentage indicator (Model 605 only): O
2
< 70%
ALERT CONDITIONS
H600/605
Concentrators have a yellow visual LED with intermittent audible alarm and red LED:
•
No oxygen flow
Oxygen percentage indicator (model 605): 70% < O
2
< 85%
M600/605; Enhanced M600/605
Concentrators have a yellow LED with intermittent audible alarm and
red LED:
•
No oxygen flow
Oxygen percentage indicator (model 605): 70% < O
2
< 82%
POWER INDICATION
Visual, green LED (next to Power Switch).
OXYGEN PERCENTAGE INDICATOR (MODEL 605)
H605 Concentrators
Warnings Conditions: Blinking yellow LED after unit is turned on and until O
2
> 85%.
Normal Oxygen: Visual green LED when O
2
> 85%.
M605; Enhanced M605 Concentrators
Warnings Conditions: Blinking yellow LED after unit is turned on and until O
2
> 82%.
Normal Oxygen: Visual green LED when O
2
> 82%.