Preparing for Use
R&S
®
SMB100A
45
Operating Manual 1407.0806.32 ─ 21
b) Deactivate the VNC Server service.
The connection is disabled, the VNC icon disappears from the task bar of the
instrument.
2.8 LXI Configuration
LXI ("LAN eXtensions for Instrumentation" is an instrumentation platform for measuring
instruments and test systems that is based on standard Ethernet technology. LXI is
intended to be the LAN-based successor to GPIB, combining the advantages of Ether-
net with the simplicity and familiarity of GPIB.
Like GPIB, LXI determines and standardizes the way the instrument behaves in a LAN.
The LXI implementation by the Rohde&Schwarz signal generators allows you to
change certain LAN settings, to reset the LAN connection as well as to identify the
instrument easily.
For information about the LXI standard, refer to the LXI website at
. See also "News from Rohde & Schwarz, article 2006/II - 190".
The R&S
SMB provides an integrated "LXI Status" dialog for LXI status indication and
reset of the LAN configuration ("LAN Configuration Initialize", LCI). To access the LXI
status dialog, press the SETUP key and select "Remote > LXI Status".
Chapter 5.2.3.16, "LXI Status"
Firmware update
After a firmware update, shut down and restart the instrument in order to enable the full
LXI functionality.
Default state of the network settings
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter
Value
TCP/IP Mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP Ping
Enabled
Password for LAN configuration
LxiWebIfc
The LCI for the R&S
SMB also resets the following parameters:
LXI Configuration