Remote Control Commands
R&S
®
SMB100A
316
Operating Manual 1407.0806.32 ─ 21
:SENSe<ch>[:POWer]:FILTer:SONCe
..............................................................................320
:SENSe<ch>[:POWer]:FILTer:TYPE
................................................................................ 320
:SENSe<ch>[:POWer]:LOGGing:STATe
...........................................................................321
..................................................................................321
........................................................................................ 322
:SENSe<ch>[:POWer]:OFFSet:STATe
............................................................................. 322
.................................................................................... 322
.......................................................................................323
SENSe<ch>[:POWer]:STATus[:DEVice]?
......................................................................... 323
....................................................................................323
..........................................................................................324
.......................................................................................... 324
............................................................................................324
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
<UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command
Parameters:
<UseDefAp>
0 | 1 | OFF | ON
*RST:
1
Example:
SENS:POW:APER:DEF:STAT 0
deactivates the default aperture time of the sensor.
Manual operation:
See
:SENSe<ch>[:POWer]:APERture:TIMe
<ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
Parameters:
<ApTime>
float
Range:
depends on connected power sensor
Increment: 1E-9
*RST:
depends on connected power sensor
Example:
SENS:POW:APER:TIM 23ms
sets 23 ms aperture time.
Manual operation:
See
:INITiate<ch>[:POWer]:CONTinuous
<Continuous>
The command switches the local state of the continuous power measurement by the
R&S NRP-Zxx power sensors on and off. Switching the local state off enhances the
measurement performance during remote control
SENSe, READ and INITiate Subsystems