Remote Control Commands
R&S
®
SMB100A
399
Operating Manual 1407.0806.32 ─ 21
Manual operation:
See
"Gate Input Polarity - Pulse Generator"
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance
<Impedance>
Selects the impedance for external pulse trigger.
Parameters:
<Impedance>
G50 | G10K
*RST:
G50
Example:
SOUR:PULM:TRIG:EXT:IMP G50
selects 50 Ohm as the trigger impedance for the external pulse
trigger.
Options:
R&S SMB-K21 or R&S SMB-K22 (Pulse Modulator)
Manual operation:
See
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe
<Slope>
Sets the polarity of the active slope of an applied trigger at the PULSE EXT connector.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Example:
PULM:TRIG:EXT:SLOP NEG
The pulse generator is triggered on the negative slope of the
external trigger signal.
Options:
R&S SMB-K23 (Pulse Generator)
Manual operation:
See
"External Trigger Input Slope - Pulse Generator"
[:SOURce<hw>]:PULM:TRIGger:MODE
<Mode>
Selects the trigger mode for pulse modulation.