Remote Control Commands
R&S
®
SMB100A
382
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:POWer:MANual
<Manual>
In Sweep mode
(:SOUR:POW:MODE SWE
) the command sets the level for the next
sweep step in the Step sweep mode (
:SOUR:SWE:POW:MODE MAN
). Here only level
values between the settings
[:SOUR]:POW:STAR
and
[:SOUR]:POW:STOP
are per-
mitted. Each sweep step is triggered by a separate
:SOUR:POW:MAN
command.
As with the "Level" value entered in the "RF Level" menu, the
OFFSet
value is also
taken into consideration with this command.
The specified value range is therefore only effective if
:SOURce:POWer:OFFSet
is set
to 0. The value range for other
OFFset
values can be calculated using the following
formula:
Minimum level +
OFFSet
... Maximum level +
OFFSet
Parameters:
<Manual>
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model
The values are given in the data sheet.
Increment: 0.01
*RST:
-30
Example:
POW:SWE:MODE MAN
sets the Step sweep mode for RF output A.
POW:MAN -5 dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode for RF output A.
POW:MODE SWE
sets the Level Sweep mode for RF output A.
POW:MAN -5.5 dBm
triggers the next sweep step with a level of -5.5 dBm.
Manual operation:
See
[:SOURce<hw>]:POWer:MODE
<Mode>
Sets the instrument operating mode and therefore also the commands used to set the
output level.
SOURce Subsystem