Instrument Function
R&S
®
SMB100A
160
Operating Manual 1407.0806.32 ─ 21
"Frequency /
Hz"
Enters the frequency to which the level correction value applies.
Note:
The "Fill..." function allows to automatically enter any number
of frequencies with freely selectable range and increment.
Using the "Fill With Sensor" function of the "Edit" sub menu requires
only the entry of the frequency values. The level values are automati-
cally acquired by the connected power sensor.
"Power/dB"
Enters the level correction value to which the specified frequency
applies. The values can be entered manually or automatically with the
"Fill With Sensor" function (available in the "Edit" sub menu).
"Goto"
Selects row for editing.
If Goto row is selected, a window opens for entering the requested
row.
"Edit"
Calls a selection of possible actions described below.
"Insert Row"
Insert a new row before the marked row.
"Insert Range"
Insert new rows before the marked row. The number of rows to be
inserted can be defined in an entry window.
"Fill...."
Opens a sub menu for defining a set of list values to be automatically
entered in the ucor list (see
"Filling the Correction List automatically"
RF Block