Remote Control Commands
R&S
®
SMB100A
336
Operating Manual 1407.0806.32 ─ 21
Example:
CORR:CSET '/var/user/ucor1'
selects the table
ucor1
.
CORR ON
activates level correction. Correction is performed using the
table
ucor1
.
Manual operation:
See
"User Cor. Data - User Correction"
[:SOURce<hw>]:CORRection[:STATe]
<State>
Activates/deactivates level correction. Level correction is performed using the table
which has been selected with the command
[:SOURce<hw>]:CORRection:CSET[:
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SOUR:CORR:CSET '/var/user/ucor1'
selects the table
ucor1
.
SOUR:CORR ON
activates user correction.
Manual operation:
See
[:SOURce<hw>]:CORRection:VALue?
Queries the current value for user correction.
Return values:
<Value>
float
Range:
-100 to 100
Increment: 0.01
*RST:
0
Example:
CORR:VAL?
queries the value currently used for level correction.
Response:
-3
the correction value is - 3 dB.
Usage:
Query only
Manual operation:
See
"User Correction Value - User Correction"
[:SOURce<hw>]:CORRection:ZERoing:STATe
<State>
Activates the zeroing procedure before filling the user correction data acquired by a
sensor.
Parameters:
<State>
0 | 1 | OFF | ON
Manual operation:
See
"Fill User Correction Data with Sensor"
SOURce Subsystem