Preparing for Use
R&S
®
SMB100A
26
Operating Manual 1407.0806.32 ─ 21
2.3.2 Unpacking and Checking the Instrument
To remove the instrument from its packaging and check the equipment for complete-
ness, proceed as follows:
1. Pull off the polyethylene protection pads from the instrument's rear feet.
2. Carefully remove the pads from the instrument handles at the front.
3. Pull off the corrugated cardboard cover that protects the rear of the instrument.
4. Carefully unthread the corrugated cardboard cover at the front that protects the
instrument handles and remove it.
5. Check the equipment for completeness using the delivery note and the accessory
lists for the various items.
6. Check the instrument for any damage. If there is damage, immediately contact the
carrier who delivered the instrument. Make sure not to discard the box and packing
material.
Packing material
Retain the original packing material. If the instrument needs to be transported or ship-
ped at a later date, you can use the material to protect the control elements and con-
nectors.
2.3.3 Placing or Mounting the Instrument
The R&S
SMB is designed for use under laboratory conditions, either on a bench top
or in a rack using a rack adapter kit (order number see data sheet).
Bench Top Operation
If the R&S
SMB is operated on a bench top, the surface should be flat. The instrument
can be used in horizontal position, standing on its feet, or with the support feet on the
bottom extended.
Putting into Operation