Remote Control Commands
R&S
®
SMB100A
346
Operating Manual 1407.0806.32 ─ 21
f
STARt
* f
MULTiplier
+ f
OFFset
... f
STOP
* f
MULTiplier
+ f
OFFset
Parameters:
<Start>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
100 MHz
Example:
FREQ:STARt 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
Manual operation:
See
"Start Freq - Frequency Sweep"
[:SOURce<hw>]:FREQuency:STOP
<Stop>
Sets the stop frequency for the RF sweep.
This parameter is related to the center frequency and span. If you change the fre-
quency, these parameters change accordingly.
f
STARt
> f
STOP
is permitted.
f
STOP
= (f
CENTer
+ f
SPAN
/2).
Note:
A defined offset affects the sweep range and consequently all correlating param-
eters. The set frequencies are only absolute values, if the Offset = 0. Offset
≠
0 shifts
the frequencies according to the offset value.
f
STARt
* f
MULTiplier
+ f
OFFSet
... f
STOP
* f
MULTiplier
+ f
OFFSet
Parameters:
<Stop>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
500 MHz
Example:
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
FREQ:STAR 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:STEP[:INCRement]
<Increment>
Sets the step width for
To adjust the frequency step by step with this step size, use the
FREQ:UP
and
FREQ:DOWN
commands.
SOURce Subsystem