Remote Control Commands
R&S
®
SMB100A
331
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of level values in the selected table.
The numerical suffix at
SOURce
must not be used for this command.
Return values:
<Points>
integer
Range:
0 to 10000
*RST:
0
Example:
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW:POIN?
queries the number of level values in the table ucor1.
Response:
440
the table ucor1 contains 440 level values.
Usage:
Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
The command fills the selected user correction list with the level values measured by
the power sensor for the given frequencies.
To select the used power sensor set the suffix in key word
SENSe
.
Example:
CORR:CSET:DATA:SENS:POW:SONC
fills the user correction list with level values acquired by the
power sensor connector to the SENSOR connector.
Usage:
Event
Manual operation:
See
"Fill User Correction Data with Sensor"
[:SOURce]:CORRection:CSET:DELete
<Filename>
Deletes the specified table.
The lists are stored with the fixed file extensions
*.uco
in a directory of the user's
choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR
. A path can also be specified in command
SOUR:CORR:CSET:CAT?
,
in which case the file in the specified directory is deleted.
Setting parameters:
<Filename>
<table name>
Example:
MMEM:CDIR '/var/user/ucor'
selects the directory for the user correction files.
CORR:CSET:DEL 'UCOR1'
deletes the table
ucor1
.
Usage:
Setting only
Manual operation:
See
"User Cor. Data - User Correction"
SOURce Subsystem