[:SOURce<hw>]:PULM:DOUBle:DELay
<Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay>
float
Range:
10 ns to 100 s
Increment: 5 ns
*RST:
3 us
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
Options:
R&S
SMB-K23 (Pulse Generator)
Manual operation:
See
"Double Pulse Delay - Pulse Generator"
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Enables/disables double pulse generation. The two pulses are generated in one pulse
period.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
Options:
R&S SMB-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:DOUBle:WIDTh
<Width>
Sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width>
float
Range:
10 ns to 100 s
Increment: 10 ns
*RST:
3 us
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Options:
R&S
SMB-K23 (Pulse Generator)
Manual operation:
See
"Double Pulse Width - Pulse Generator"
[:SOURce<hw>]:PULM:MODE
<Mode>
Sets the mode of the pulse generator.