Instrument Function
R&S
®
SMB100A
184
Operating Manual 1407.0806.32 ─ 21
"Step"
Sets a step-by-step sweep cycle.
If this mode is activated, the cursor moves to the value displayed for
"Current Level". Each sweep step is triggered by a variation of the
value in the "Current Level" entry window. The step width is set below
at entry field "Step".
If this mode is activated, the cursor moves to the value displayed for
"Current Level". If a different sweep mode was activated prior to the
"Step" mode, the current sweep is stopped. The step sweep starts at
the current level value.
Example:
SOUR:SWE:POW:MODE MAN
TRI:PSW:SOUR SING
SOUR:SWE:POW:STEP 0.5
SOUR:POW:MODE SWE
SOUR:POW:MAN -16
The value entered with command
SOUR:SWE:POW:STEP
sets the
step width.
The value entered with command
SOUR:POW:MAN
has no effect, the
command only triggers the next sweep step. However, the value has
to be in the currently set sweep range (start to stop). In remote con-
trol only a step-by-step sweep from start to stop frequency is possi-
ble.
"Extern Single" Sets a single sweep cycle. The sweep is triggered by an external trig-
ger signal.
Refer to the description of the rear panel for information about the
connectors for external trigger signal input (see
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EXT
SOUR:POW:MODE SWE
(External trigger)
"Extern Step"
Sets a step-by-step sweep cycle. Each sweep step is triggered by an
external trigger signal (trigger source as described under "Extern Sin-
gle"). The step width corresponds to the step width of the rotary knob.
Example:
SOUR:SWE:POW:MODE STEP
SOUR:SWE:POW:STEP 0.5
TRIG:PSW:SOUR EXT
SOUR:POW:MODE SWE
(External trigger)
RF Block