Remote Control Commands
R&S
®
SMB100A
345
Operating Manual 1407.0806.32 ─ 21
Parameters:
<Offset>
float
Increment: 0.01
*RST:
0
Example:
FREQ:OFFS 500kHz
sets the frequency offset to 500 kHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:SPAN
<Span>
Determines the extent of the frequency sweep range. This setting in combination with
the center frequency setting (
[:SOURce<hw>]:FREQuency:CENTer
) defines the
sweep range.
This parameter is related to the start and stop frequencies. If you change the fre-
quency, the span changes accordingly.
f
SPAN
= f
STOP
- f
STARt
f
STARt
> f
STOP
is permitted.
Parameters:
<Span>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
400E6
Example:
FREQ:CENT 400 MHz
sets the center frequency of the frequency sweep to 400 MHz.
FREQ:SPAN 200 MHz
sets a span of 200 MHz. This sets the sweep range to 300 MHz
to 500 MHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:STARt
<Start>
Sets the start frequency for the RF sweep.
This parameter relates to the center frequency and span. If you change the frequency,
these parameters change accordingly.
f
STARt
> f
STOP
is permitted.
f
STARt
= (f
CENTer
- f
SPAN
/2).
Note:
A defined offset and the multiplier factor affect the sweep range and therefore all
correlated parameters. The set frequencies are only absolute values, if the offset = 0
and the multiplication factor = 1. The multiplier multiplies the frequencies accordingly,
and the offset ≠ 0 shifts the frequencies corresponding to the set value.
SOURce Subsystem