Remote Control Commands
R&S
®
SMB100A
373
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:PHASe:REFerence
Adopts the phase set with
SOURce:PHASe:ADJust
as the current phase.
Example:
PHAS 0.1RAD
changes the phase by 0.1 RAD relative to the current phase.
PHAS:REF
adopts the set phase as the current phase.
Usage:
Event
Manual operation:
See
7.13.11 SOURce:PM Subsystem
The PM subsystem contains the commands for checking the phase modulation. The
settings for the internal modulation source (LF generator) are made in the
For information on the required options, see
Chapter 5.4.4, "Phase Modulation (PhiM)"
...................................................................................... 373
[:SOURce<hw>]:PM:EXTernal:COUPling
......................................................................... 373
[:SOURce<hw>]:PM:EXTernal:DEViation
......................................................................... 374
[:SOURce<hw>]:PM:INTernal:DEViation
.......................................................................... 374
............................................................................................ 375
.................................................................................... 375
......................................................................................... 375
............................................................................................ 376
[:SOURce<hw>]:PM[:DEViation]
<Deviation>
Sets the deviation of the phase modulation signals in RAD. The maximum deviation
depends on the set RF frequency and the selected modulation mode (see data sheet).
Parameters:
<Deviation>
float
Range:
see data sheet
Increment: 1E-6
Example:
PM 2
sets 2 RAD deviation to the phase modulation signal.
Manual operation:
See
[:SOURce<hw>]:PM:EXTernal:COUPling
<Coupling>
Selects the coupling mode for the external phase modulation signal.
SOURce Subsystem