Instrument Function
R&S
®
SMB100A
189
Operating Manual 1407.0806.32 ─ 21
General Settings
State - List Mode
Activates/deactivates the List mode. The currently selected list is processed.
In case of a new or modified list, the necessary hardware settings are automatically
determined on activation of the list mode. The data determined in this way is stored
along with the list and is available whenever the list is used again.
This means that when activating the list mode, the system checks whether any hard-
ware settings are present. If so, the list is started immediately, but if not they are auto-
matically determined (the list is learnt).
A "Learn List Mode Data" button is available for deliberately activating list learning.
Note:
Activating the list mode automatically deactivates all sweeps. During list mode
the frequency and level indications do not display the currently set values.
Remote command:
Mode - List Mode
Selects the cycle mode of the List mode.
"Auto"
Cycle from the beginning to the end of the list with automatic restart
at the beginning. If a different mode was activated prior to the Auto
mode, the cycle continues from the beginning of the list. The duration
of a list step is determined by the set dwell time.
Button "Reset" restarts the list at the starting point.
"Single"
Single cycle from the beginning to the end of the list. If "Single" is
selected, the cycle is not started immediately. The "Execute Single"
button appears under the "Mode" line. The cycle is started with this
button. The duration of a list step is determined by the set dwell time.
Button "Reset" restarts the list at the starting point.
RF Block