Remote Control Commands
R&S
®
SMB100A
445
Operating Manual 1407.0806.32 ─ 21
:SYSTem:SHUTdown
Shuts down the instrument.
Usage:
Event
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:STAR:COMP
Response: 1
The startup of the instrument is completed.
Usage:
Query only
:SYSTem:DISPlay:UPDate
<Update>
Switches the update of the display on/off. A switchover from remote control to manual
control always sets the status of the update of the display to ON.
Parameters:
<Update>
0 | 1 | OFF | ON
*RST:
ON
Example:
SYST:DISP:UPD OFF
switches update of displayed parameter values off.
Manual operation:
See
:SYSTem:DATE
<Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
This parameter is protected, in order to prevent accidental changes.
It can be accessed with protection level 1, see
Parameters:
<Year>
<year>,<month>,<day>
<Month>
integer
Range:
1 to 12
<Day>
integer
Range:
1 to 31
SYSTem Subsystem