Remote Control Commands
R&S
®
SMB100A
320
Operating Manual 1407.0806.32 ─ 21
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
<MTime>
The command defines the timeout for fixed noise filter mode
(
:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio
). This value ensures limited
settling times.
Parameters:
<MTime>
float
Range:
1 to 999.99
Increment: 0.01
*RST:
4
Default unit: s
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR .2
sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
limits the settling time to 5 seconds
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:SONCe
The command activates the search for the optimum filter length for the current mea-
surement conditions. The found filter length can be retrieved with com-
mand
:SENSe:POWer:FILTer:LENGth:USER?
. This command is only available for
user filter mode (
:SENSe:POWer:FILTer:TYPE USER
).
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:SONC
activates the search for the optimum filter length.
SENS:FILT:LENG?
returns the found optimum filter length.
Response: 128
Usage:
Event
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:TYPE
<Type>
The command selects the filter mode. The filter length is the multiplier for the time win-
dow and thus directly influences the measurement time.
SENSe, READ and INITiate Subsystems