Remote Control Commands
R&S
®
SMB100A
377
Operating Manual 1407.0806.32 ─ 21
...........................................................................................381
..................................................................................... 381
.....................................................................................382
.......................................................................................382
......................................................................................383
[:SOURce<hw>]:POWer:SPC:CRANge
............................................................................ 383
............................................................................... 384
................................................................................ 384
[:SOURce<hw>]:POWer:SPC:SELect
.............................................................................. 384
...............................................................................385
[:SOURce<hw>]:POWer:SPC:TARGet
............................................................................. 385
.......................................................................................385
[:SOURce<hw>]:POWer:STEP[:INCRement]
.................................................................... 386
............................................................................. 386
........................................................................................387
[:SOURce<hw>]:POWer:ALC:OMODe
<OffMode>
The command sets the level control mode which becomes active when automatic level
control is deactivated (ALC Off).
Parameters:
<OffMode>
SHOLd
SHOLd
Level control is activated briefly if the level or frequency changes
("ALC Off Sample & Hold").
*RST:
SHOLd
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:OMOD SHOL
level control is briefly activated if the frequency or level changes.
[:SOURce<hw>]:POWer:ALC:SONCe
Temporarily activates level control for correction purposes.
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:SONC
level control is performed once only.
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
Activates/deactivates automatic level control.
SOURce Subsystem