Remote Control Commands
R&S
®
SMB100A
386
Operating Manual 1407.0806.32 ─ 21
Parameters:
<Start>
float
Determines the first level value of the sweep setting range.
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
-30
Example:
POW:STAR -20 dBm
sets the start level for the level sweep to -15 dBm for RF output
A.
Manual operation:
See
[:SOURce<hw>]:POWer:STEP[:INCRement]
<Increment>
Sets the step width for
To adjust the level step by step with this step size, use the
POW:UP
and
POW:UP
com-
mands.
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the level accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
float
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
1
Example:
POW:STEP 2
sets the step width for entering the RF level to 2 dB.
Manual operation:
See
[:SOURce<hw>]:POWer:STEP:MODE
<Mode>
Activates (
USER
) or deactivates (
DECimal
) the user-defined step width used when
varying the level value with the level values
UP/DOWN
. The command is linked to set-
ting "Variation Active" for manual control, i.e. the command also activates/deactivates
the user-defined step width used when varying the level value with the rotary knob.
Parameters:
<Mode>
DECimal | USER
*RST:
DECimal
SOURce Subsystem