Remote Control Commands
R&S
®
SMB100A
418
Operating Manual 1407.0806.32 ─ 21
[SOURce:]SWEep[:FREQuency]:EXECute
It is recommended that you switch off the "Start/Stop Display Update" for optimum
sweep performance, especially with short dwell times (
SYSTem:DISPlay:UPDate
OFF
).
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
..................................................................418
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
.............................................................. 419
[:SOURce<hw>]:SWEep[:FREQuency]:LFConnector
......................................................... 419
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
.................................................................. 419
[:SOURce<hw>]:SWEep[:FREQuency]:OVOLtage:STARt
.................................................. 420
[:SOURce<hw>]:SWEep[:FREQuency]:OVOLtage:STOP
...................................................421
[:SOURce<hw>]:SWEep[:FREQuency]:POINts
................................................................. 421
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace
..............................................................422
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?
............................................................ 422
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe
................................................................. 422
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
...............................................................423
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
.......................................................423
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
................................................ 424
[:SOURce<hw>]:SWEep:POWer:DWELl
.......................................................................... 425
[:SOURce<hw>]:SWEep:POWer:EXECute
....................................................................... 425
[:SOURce<hw>]:SWEep:POWer:MODE
...........................................................................425
[:SOURce<hw>]:SWEep:POWer:POINts
.......................................................................... 426
[:SOURce<hw>]:SWEep:POWer:RETRace
.......................................................................427
[:SOURce<hw>]:SWEep:POWer:RUNNing?
..................................................................... 427
[:SOURce<hw>]:SWEep:POWer:SHAPe
..........................................................................427
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
........................................................... 428
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]
....................................................... 428
[:SOURce<hw>]:SWEep:RESet[:ALL]
.............................................................................. 429
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
<Dwell>
Sets the time taken for each frequency step of the sweep.
The keyword [
:FREQuency
] can be omitted (see example). The command is then
SCPI-compliant.
Tip:
It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (
Parameters:
<Dwell>
float
Range:
2E-3 to 100
Increment: 100E-6
*RST:
15E-3
Example:
SWE:DWEL 12 ms
sets a dwell time of 12 ms for a frequency sweep at the RF out-
put.
Manual operation:
See
"Dwell Time - Frequency Sweep"
SOURce Subsystem