Remote Control Commands
R&S
®
SMB100A
419
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Starts an RF frequency sweep cycle manually.
The command is only effective in single mode.
Example:
TRIG:FSW:SOUR SING
SOUR:SWE:FREQ:MODE AUT
SWE:FREQ:EXEC
triggers a frequency sweep at the RF output.
Usage:
Event
Manual operation:
See
"Execute Single Sweep - Frequency Sweep"
[:SOURce<hw>]:SWEep[:FREQuency]:LFConnector
<LfConnector>
Activates the output of a sweep voltage ramp at the LF connector.
The voltage range is set with commands
SOURce:SWEep:FREQuency:OVOLtage:STARt
and
…:STOP
Parameters:
<LfConnector>
0 | 1 | OFF | ON
*RST:
0
Example:
SWE:LFC ON
activates the output of a linear voltage ramp from sweep start to
sweep stop at the LF connector.
SWE:OVOL:STAR 0V
SWE:OVOL:STOP 3V
'the voltage at sweep start is 0 Volt and at sweep stop 3 V.
Manual operation:
See
"Use LF connector to output sweep voltage - RF Frequency
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
<Mode>
Sets the sweep mode.
The keyword
[:FREQuency]
can be omitted (see example). The command is then
SCPI-compliant.