Remote Control Commands
R&S
®
SMB100A
385
Operating Manual 1407.0806.32 ─ 21
Example:
POW:SPC:SEL SENS2
selects the sensor connected to a second USB interface for
power control.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:STATe
<PowCntrlState>
Activates power control using the selected sensor. The control loop periodically adjusts
the generator output. After switching off, the running loop is completed.
Parameters:
<PowCntrlState>
0 | 1 | OFF | ON
*RST:
0
Example:
POW:SPC:STAT ON
activates power control.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:TARGet
<PowCntrlTarget>
Sets the nominal level expected at the input of the sensor. To define the unit of the
power value, use command
Parameters:
<PowCntrlTarget>
float
Range:
-50 to 30
Increment: 0.01
*RST:
-10
Example:
SENS:UNIT dBm
selects unit dBm for setting the target level value.
POW:SPC:TARG -10
sets -10 dBm target level.
Manual operation:
See
[:SOURce<hw>]:POWer:STARt
<Start>
Sets the start level for the RF sweep.
Note:
You can select any level within the setting range. The range is defined by this
start value and the
A defined offset (
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
affects the level values according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
SOURce Subsystem