Remote Control Commands
R&S
®
SMB100A
347
Operating Manual 1407.0806.32 ─ 21
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the frequency accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
1E6
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:STEP:MODE
<Mode>
Activates (
USER
) or deactivates (
DECimal
) the user-defined step width used when
varying the frequency value with the frequency values
UP/DOWN
. The command is
linked to the command "Variation Active" for manual control, i.e. the command also
activates/deactivates the user-defined step width used when varying the frequency
value with the rotary knob.
Parameters:
<Mode>
DECimal | USER
*RST:
DECimal
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
FREQ:STEP:MODE USER
actives this step width for frequency variation with the rotary
knob (manual control) and with frequency values
UP/DOWN
(remote control).
Manual operation:
See
7.13.5 SOURce:INPut Subsystem
The
SOURce:INPut
subsystem contains the commands for configuring the inputs for
external modulation signals. The instrument trigger setting influences all sweeps and is
effective in the List mode (Instrument Trigger).
[:SOURce<hw>]:INPut:MODext:IMPedance
......................................................................347
[:SOURce<hw>]:INPut:MODext:WIGNore
.........................................................................348
................................................................................... 348
[:SOURce<hw>]:INPut:MODext:IMPedance
<Impedance>
Sets the impedance for an externally applied modulation signal.
SOURce Subsystem