Remote Control Commands
R&S
®
SMB100A
444
Operating Manual 1407.0806.32 ─ 21
Note
: While working in a emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command
SYSTem:LANGuage
will be discarded.
The return to the SCPI command set of the R&S
SMB can only be performed by using
the appropriate command of the selected command set. For example, the HP com-
mand
EX
returns to the instrument-specific GPIB command set (selection
SYST:LANG
'HPxxxx'
).
Parameters:
<Language>
string
Example:
SYSTem:LANGuage "SCPI"
sets the SCPI command set.
Manual operation:
See
:SYSTem:PROTect<ch>[:STATe]
<State>[, <Key>]
Activates/deactivates the specified protection level.
Parameters:
<State>
select
*RST:
1
Setting parameters:
<Key>
integer
The respective functions are disabled when the protection level
is activated. No password is required for activation. A password
must be entered to deactivate the protection level. The pass-
word for the first level is 123456.
This protection level can be used to lock-out internal adjust-
ments.
Example:
SYSTem:PROTect1:STATe ON
activates protection level 1. Internal adjustments are only possi-
ble after deactivating the lock-out.
SYSTem:PROTect1:STATe OFF,123456
deactivates protection level 1. Internal adjustments are enabled
again.
Manual operation:
See
:SYSTem:REBoot
Restarts the firmware and the operating system.
Usage:
Event
:SYSTem:RESTart
Restarts the firmware. The operating system remains active.
Usage:
Event
SYSTem Subsystem