VIBSCANNER 11.2010
70
Technical data
Hardware
Measurement channels
Analog: Vibration signals (LineDrive, ICP*)
Temperature (Pt 100*, NiCrNi)
Transducer and instrument outputs
AC (± 30V; 0-20mA*)
DC (± 30V; 0-20mA*)
Digital:
Trigger (5V TTL)
Outputs
RS 232 (PC conn., up to 115 kbaud), Head-
phone, Analog signal (4V
pp
; R
out
= 200 Ohm)
Operating elements
1 joystick (Cursor & ENTER function)
2 keys
(Menu and Escape)
LED indicators
Four LEDs for device status / signal evaluation
Display
Graphical pixel display (backgr. illumination)
Dimensions
54 x 27mm/ 2.1 x 1.1inch
128 x 64 pixels
Contrast
Adjustable
Illumination
Adjustable
Power supply
NiMH rechargeable pack (7.2V / 1.5 Ah)
Charg. dur.
< 6 hours (intr. safe: <10 hours)
Operat. dur.
> 10 hours in intermit. use
> 6 hours in continuous use
with illumination
Charg. display 2 LEDs (green, red)
Charg. temp. 10 ... 40°C / 50 ... 104°F
‘Sleep’ mode Adjustable
Internal sensors
Vibration/ shock pulse (bearing condition)
Frequency range: 10Hz ... 10kHz** (±10%)
Resonance frequency: 36 kHz**
RPM (IR sensor with light point for adjustment)
Temperature (NiCrNi)
Signal processing
RMS, 0-P, P-P, max/carpet, envelope curve,
rectifying
Filter:
Highpass (2 /10 Hz; 1 / 5 kHz)
Lowpass (1 / 5 / 10***/ 40 kHz)
Integrat.: Two selectable stages
Sampling frequencies: Up to 64kHz (depending
on measuring range)
Memory
512 MB (intr. safe: 4 MB)
Housing
ABS strengthened with steel fiber
Environmental protection
IP 65
Relative humidity
10 ... 90%
Temperature range
Operation
0... +60°C / 32...140°F
Storage
-20... +80°C / -4...176°F
Dimensions (HxWxD)
250 x 100 x 55 mm / 9.8 x 3.9 x 2.2 inch
Weight
approx. 690 g / 24 oz.
Transport case
Polypropylene, black
Dimensions
390 x 340 x 90 mm
(HxWxD)
15.3 x 13.4 x 3.5 inch
Measurement range / Accuracy
RPM
60 ... 60000 min
-1
/ 0.1‰
Temperature
• Pt 100
-50..+600°C / 1°+ sensor%
(-58 ...+1112°F)
• NiCrNi
-50...+100°C / 0.5° + 3%
(internal)
(-58...+212°F)
(external)
-50...+100°C / 0.5°+ sensor%
(-58...+212°F)
(external)
+100..+1000°C / 1°+ sensor%
(212...1832°F)
Extra-low
-9...+9V / 2% (R
i
=30kOhm,
voltage
with cable VIB 5.440)
(AC/DC)
-30...+30V / 2% (R
i
=100kOhm,
with cable VIB 5.433)
Extra-low
-20...+20mA / 2%;
current
4...20mA / 2% (R
shunt
=200 Ohm
(AC/DC)
with cable VIB 5.434)
For internal sensors, external sensors (1µA/ms
-
²
CLD*; 100mV/g ICP
®
) and external measure-
ment devices (1mV/ms
-
²), the following applies:
Displacement < 9000 µm (p-p) / 1%
Velocity
< 9000 mm/s (p-p) / 1%
Acceleration
< 961 m/s² (p-p) / 1%
Shock pulses < 81 dBsv / ± 3dB
Fulfilled standards
Frequency response according to ISO 2954 -
other parameters and measured variables
according to DIN 45662 class 1
Noise, internal sensor (from 10 Hz)
Velocity
0.1 mm/s eff.
Displacement 2µm eff. (instr.+sensor)
Shock pulse
< 0dBsv , peak
Appendix
* not for VIBSCANNER EX,
** in cone sinking,
*** optional for VIBSCANNER EX