VIBSCANNER 11.2010
22
R
EGISTRATION
: Click on ‘Select’ to register the optional
firmware modules (FFT, Balance, Analysis). Click on the
corresponding module and enter the password in the
text editor. The ‘FFT’ and ‘Analysis’ modules can be
enabled for a trial period of 30 hours of operation.
Communication with the OMNITREND PC software can
be enabled via ‘PC License VIB.’
D
EVICE
INFO
: Click on ‘Display’ to show the following infor-
mation about the instrument:
- Serial number, ID number
- Date of the next calibration; A follow-up calibration is
required two years after the instrument is delivered. An
appropriate warning message first appears on the dis-
play four weeks before the deadline.
- Available memory in percent*.
- Hardware status: instrument/ digital and analog board,
sensor carrier module, joystick board, LED board.
This information is required in the case of a repair or
calibration.
O
FFSET
C
OMP
.: Various factors (aging, temperature) cause an
offset in the DC circuit that increases with time. Thus,
this offset should be balanced approximately every 2
months. Click on ‘Select’ to start the compensation
procedure.
A
UTO
O
RBIT
: Only applies for orbit measurements with the
‘Signal analysis’ module. Recording of the shaft move-
ment begins automatically after the RPM measurement
(‘Yes’).
S
IGNAL
OUTPUT
: Setting the signal output (yellow socket) to
100 mV/g, ‘headphone’ or 1 mV/ms-² (see page 51).
* Max. 97-99%, as 1-3% are
occupied by languages,
transducers and evaluation.
Getting started