VIBSCANNER 11.2010
33
Multimeter
User-defined measurement tasks
Process parameters (e.g. pressure) can be measured as a
level (DC) or as a characteristic value of a signal (AC).
Level measurement (DC)
To check and set up the measuring parameters:
• Select the icon for level measurements in the main
window.
• Open the ‘Measurement’ setup (Setup -> Measurement)
and set up the number of averages, the delay and the
type, if necessary.
• Open the ‘Transducer’ setup (Setup -> Transducer) and
set up the following parameters:
-> (F
IRST
L
INE
):
Opens the transducer list
M
EAS
.T
YPE
:
User-defined quantity
S
IGNAL
TYPE
:
DC voltage or DC current
E
DIT
QUANTITY
:
Description of the measured quantity
E
NGIN
. U
NIT
::
Unit as ‘engineering unit’
D
ECIMAL
:
Decimal places of the results display
S
ENSITIVITY
:
Sensitivity of the transducer
O
FFSET
:
Offset of the transducer
To perform the measurement, connect the VIBSCANNER to
the transducer and click on the relevant icon in the main
window.
Signal measurement (AC)
For a user-defined signal measurement (AC), proceed in
the same way as described in the previous section. The
frequency range and resonating frequency must be entered
as additional transducer parameters.