Maxprep™ Liquid Handler | TM509 | Revised 1/19
Operating the Maxprep™ Liquid Handler
5.3 Starting a Method (continued)
Figure 31. Selecting a method to run from the ‘Methods’ screen. Select a method or
administrator-defined variant method from the ‘Methods’ screen. Once a method is selected
it will be highlighted in yellow. Press the
Proceed button to start the method.
Alternative method selection for preprocessing methods
To select an appropriate preprocessing method for a Maxwell
®
RSC purification kit, touch the
text box at the top of the ‘Methods’ screen (Figure 31) and either scan or type the bar code on
the Maxwell
®
RSC reagent kit box indicated by the Scan Here label (Figure 32). The Maxprep™
software requires manual entry of bar codes in the following format: Product Catalog Number,
Kit Lot Number, Expiration Date in year-month format (example: AS13303221872020-05, where
the product catalog number is AS1330, the kit lot number is 322187 and the expiration date is
2020-05). Only if the bar code entered has this format will a method be selected. If applicable,
select any of the other variants that you wish to run to change the selection.
There are two possible warnings that can be displayed based on entry of kit bar code
information. For kits that have passed the expiration date, the Maxprep™ software will display
a warning that the kit is past expiration. If the administrator has allowed the use of expired kits,
the user can make the choice of whether to proceed with this method or not; otherwise, the
software will not allow the selection of a method based on a bar code that indicates the kit has
expired. When there is no corresponding method for a kit based on the bar code entry, the
software will display a warning of the issue. Check the Promega web site (
www.promega.com/
resources/tools/maxprepmethod
) to see if a method has been released for the kit.
14657T
A
Summary of Contents for Maxprep AS9100
Page 2: ......
Page 6: ...4 Maxprep Liquid Handler TM509 Revised 1 19 Table of Contents...
Page 16: ...14 Maxprep Liquid Handler TM509 Revised 1 19 Introduction Notes...
Page 24: ...22 Maxprep Liquid Handler TM509 Revised 1 19 Hardware Overview Notes...
Page 54: ...52 Maxprep Liquid Handler TM509 Revised 1 19 Maxprep Software Notes...
Page 74: ...72 Maxprep Liquid Handler TM509 Revised 1 19 Operating the Maxprep Liquid Handler Notes...
Page 78: ...76 Maxprep Liquid Handler TM509 Revised 1 19 Cleaning the Maxprep Liquid Handler Notes...
Page 82: ...80 Maxprep Liquid Handler TM509 Revised 1 19 Troubleshooting Notes...
Page 84: ...82 Maxprep Liquid Handler TM509 Revised 1 19 Instrument Service Notes...
Page 90: ...88 Maxprep Liquid Handler TM509 Revised 1 19 Certificate of Decontamination Notes...