7
Maxprep™ Liquid Handler | TM509 | Revised 1/19
Introduction
1.5 Maxprep™ Liquid Handler Specifications
Instrument Specifications
Desktop PC
Windows
®
10 Operating System with Ethernet connection to
instrument, wired connectivity to local network
Communication
Ethernet
Weight
Approximately 220lb (98.6kg)
Dimensions (W × D × H)
42.1 × 27.8 × 33 inches (106.9 × 70.6 × 83.3cm)
height with door open: 48.4 inches (123cm)
Power Requirements
Input: 100–240VAC, 50/60Hz, 5A
Output: 42VDC ± 5%, 600W max
250VAC, 5A, low breaking capacity, time-lag fuse (AC250V,
T5AL, 5 × 20mm)
Operating
Temperature
Relative Humidity
Altitude
15–35°C (59–95°F)
30% to 85% non-condensing
0–2,000m above sea level
Storage
Temperature
Relative Humidity
–20° to 70°C
10% to 90% non-condensing
UV Bulb (optional)
Average lifetime approximately 10,000 hours, length 9.2in
(233mm base face to face, diameter 15mm, 11W, 0.43A
current, 25.6V, UV-C, UV output 32W at 1 meter)
Regulatory
Compliance
For Research Use Only. Not for use in Diagnostic Procedures.
CE Compliance
Pollution degree 2.
Installation category II.
Indoor use only.
Main supply voltage fluctuations are not to exceed >10% of the
nominal supply voltage.
This instrument complies with the requirements of the EU
Directives for electromagnetic compatibility (EMC Directive)
and for electrical safety (LVD). Compliance with the relevant
directives was demonstrated by third-party testing to the
appropriate EN and IEC standards. The product bears the
CE mark on its label as evidence of compliance, and an EU
Declaration of Conformity for the product is available from
Promega on request.
Summary of Contents for Maxprep AS9100
Page 2: ......
Page 6: ...4 Maxprep Liquid Handler TM509 Revised 1 19 Table of Contents...
Page 16: ...14 Maxprep Liquid Handler TM509 Revised 1 19 Introduction Notes...
Page 24: ...22 Maxprep Liquid Handler TM509 Revised 1 19 Hardware Overview Notes...
Page 54: ...52 Maxprep Liquid Handler TM509 Revised 1 19 Maxprep Software Notes...
Page 74: ...72 Maxprep Liquid Handler TM509 Revised 1 19 Operating the Maxprep Liquid Handler Notes...
Page 78: ...76 Maxprep Liquid Handler TM509 Revised 1 19 Cleaning the Maxprep Liquid Handler Notes...
Page 82: ...80 Maxprep Liquid Handler TM509 Revised 1 19 Troubleshooting Notes...
Page 84: ...82 Maxprep Liquid Handler TM509 Revised 1 19 Instrument Service Notes...
Page 90: ...88 Maxprep Liquid Handler TM509 Revised 1 19 Certificate of Decontamination Notes...