75
Maxprep™ Liquid Handler | TM509 | Revised 1/19
Cleaning the Maxprep™ Liquid Handler
6.2 Dealing with Spills
Wipe up any spills immediately. If any material spills in the instrument, wipe up the material with
a cloth dampened with 70% ethanol. Note that the reagents contain hazardous materials;
therefore, dispose of the cleaning materials according to your institutional guidelines. Wipe well
once visible material is removed. In case of spills in the instrument where there is a potential
biohazard, wipe up the spill with paper towels and wash the spill area with a detergent solution
such as Steris LpH
®
following the manufacturer’s instructions. Dispose of towels used according
to your institutional guidelines for biohazardous waste.
Bleach reacts
with guanidine
thiocyanate, which
may be present
in the Maxwell
®
RSC Cartridges,
and should not be
added to any sample
waste containing
lysis solutions. Do
not use bleach to
clean Maxwell
®
RSC
reagent spills.
Summary of Contents for Maxprep AS9100
Page 2: ......
Page 6: ...4 Maxprep Liquid Handler TM509 Revised 1 19 Table of Contents...
Page 16: ...14 Maxprep Liquid Handler TM509 Revised 1 19 Introduction Notes...
Page 24: ...22 Maxprep Liquid Handler TM509 Revised 1 19 Hardware Overview Notes...
Page 54: ...52 Maxprep Liquid Handler TM509 Revised 1 19 Maxprep Software Notes...
Page 74: ...72 Maxprep Liquid Handler TM509 Revised 1 19 Operating the Maxprep Liquid Handler Notes...
Page 78: ...76 Maxprep Liquid Handler TM509 Revised 1 19 Cleaning the Maxprep Liquid Handler Notes...
Page 82: ...80 Maxprep Liquid Handler TM509 Revised 1 19 Troubleshooting Notes...
Page 84: ...82 Maxprep Liquid Handler TM509 Revised 1 19 Instrument Service Notes...
Page 90: ...88 Maxprep Liquid Handler TM509 Revised 1 19 Certificate of Decontamination Notes...