13
Maxprep™ Liquid Handler | TM509 | Revised 1/19
Introduction
1.9 Environmental Requirements (Operating, Shipping and
Storage Conditions)
Power
Requirements
100–240VAC, 50/60Hz, 5A
Temperature
–20 to 70°C (shipping/storage), 15–35°C (operation)
Humidity
30–85% relative humidity, non-condensing (operation)
10–90% non-condensing (storage)
Operating Altitude
<2000 meters
The Maxprep™ Liquid Handler is intended for indoor use only. Wipe up spills immediately. To
avoid shortening the expected lifespan of the instrument, install in a location that meets the
following criteria:
• Locate on a sturdy, level surface capable of stably supporting 350lb.
• Avoid dusty areas.
• Choose a location that has good air circulation and is not exposed to direct sunlight.
• Avoid electrically noisy power sources (e.g., power generators).
• Do not install in a location where there is large temperature variability or high humidity.
• Do not position the instrument so that it is difficult to unplug from the power source.
• Do not place next to heat sources.
• Do not use near flammable gases or liquids.
• Allow at least 6 inches (15.25cm) clearance on all sides of the Maxprep™ Liquid Handler.
• Do not place near other electrically sensitive instruments. Allow at least 6 inches (15.25cm)
on all sides of the Maxprep™ Liquid Handler.
• Connect the Maxprep™ Liquid Handler and PC to a power outlet. Promega recommends
that you plug the instrument and PC into an uninterruptible power supply.
Summary of Contents for Maxprep AS9100
Page 2: ......
Page 6: ...4 Maxprep Liquid Handler TM509 Revised 1 19 Table of Contents...
Page 16: ...14 Maxprep Liquid Handler TM509 Revised 1 19 Introduction Notes...
Page 24: ...22 Maxprep Liquid Handler TM509 Revised 1 19 Hardware Overview Notes...
Page 54: ...52 Maxprep Liquid Handler TM509 Revised 1 19 Maxprep Software Notes...
Page 74: ...72 Maxprep Liquid Handler TM509 Revised 1 19 Operating the Maxprep Liquid Handler Notes...
Page 78: ...76 Maxprep Liquid Handler TM509 Revised 1 19 Cleaning the Maxprep Liquid Handler Notes...
Page 82: ...80 Maxprep Liquid Handler TM509 Revised 1 19 Troubleshooting Notes...
Page 84: ...82 Maxprep Liquid Handler TM509 Revised 1 19 Instrument Service Notes...
Page 90: ...88 Maxprep Liquid Handler TM509 Revised 1 19 Certificate of Decontamination Notes...