![PFU Limited fi-6130 Maintenance Manual Download Page 59](http://html1.mh-extra.com/html/pfu-limited/fi-6130/fi-6130_maintenance-manual_1535731059.webp)
08
July 27, 09 K.Okada
A.Miyoshi I.Fujioka
Refer to Revision Record on page 2.
TITLE
fi-6140/fi-6240/fi-6130/fi-6230/fi-614PR
Maintenance Manual
07
Nov.12, 08 K.Okada
T.Anzai
I.Fujioka
Refer to Revision Record on page 2.
CUST.
06
Mar.10, 08 K.Okada
T.Anzai
I.Fujioka
Refer to Revision Record on page 2.
DRAW
No.
P1PA03540-B0XX/6
Rev. DATE DESIGN CHECK APPR. DESCRIPTION
Design
July 27, 2007
K.Okada
CHECK
K.Okada
APPR.
T.Anzai
PFU LIMITED
Page
59
/
257
Section 4-3-17
4-3-17 Frequent “U0: Shipping lock error” or “E0: Carrier drive alarm”
Table 4-3-17
Item
No.
Check items
How/where to check
1
Does the same symptom occur after
turning OFF and ON the scanner?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
Does the same symptom occur after
unlocking the Shipping lock?
Refer to Section 2-2-2.
3
Is there abnormal noise from the FB
Motor when this error occurs?
If abnormal noise is heard, go to item No.4.
If there is no abnormal noise, go to item No.5 to check sensors
or drive unit.
4
Does any foreign object get inside drive
unit or carrier unit?
Check visually.
5
Is the Joint PCA connected correctly?
Refer to Section 5-12-10 for removal.
6
Replace the Joint PCA and see if the
error is resolved.
Refer to Section 5-12-10.
7
Is the CR Belt tension loose?
Refer to a note of (2) of Section 5-5 for checking. Loose belt
should be tightened.
8
Is the FB Motor cable connected?
Connect it if disconnected.
9
Replace the Control PCA and see if the
error is resolved.
Refer to Section 5-11.
4-3-18 “E1/E2/E3: Optical alarm”
Table 4-3-18
Item
No.
Check items
How/where to check
1
Does the same symptom occur after
turning OFF and ON the scanner?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
When E2 or E3 error occurs, check if
the glass in the reading position dirty
Open the ADF and clean the glass. (Refer to Section 7-2-1.)
3
Is the Lamp ON?
Are the cables between the Lamp and
Inverter damaged?
Are the connectors connected correctly?
Open the ADF and press the Cover open sensor to light the
Lamp. If the Lamp does not light, replace the Lamp, and then
Inverter.
E1 (FB): Refer to Sections 5-12-7, 5-12-6
E2 (ADF front): Refer to Sections 5-9-4, 5-9-3
E3 (ADF back): Refer to Section 5-10-6, 5-10-5.
Refer to Section 6-2 “Saving EEPROM data” for the Cover
Open sensor installation position.
4
Is the OPT Unit dirty?
Is the CCD cable damaged?
Are the connectors connected correctly?
E1 (FB): Refer to Section 5-3-1 for cleaning/checking.
Referring to Section 5-12-8, check if the CR cable is damaged.
E2 (ADF front): Refer to Section 5-3-1 for cleaning/checking.
E3 (ADF back): Refer to Section 5-3-1 for cleaning/checking.
5
When the error is E2 (ADF front), check
if the glass and white reference sheet
inside of the Base Unit are dirty.
White reference sheet inside of the Base Unit: Section 5-3-5.
6
When the error is E3 (ADF back), check
if the glass and white reference sheet
inside of the Upper Unit are dirty.
White reference sheet inside of the Upper Unit: Section 5-3-6.
7
Replace OPT Unit and see if the error is
resolved.
E1 (FB): Refer to Section 5-12-5.
E2 (ADF front): Refer to Section 5-9-6.
E3 (ADF back): Refer to Section 5-10-4.
8
Replace the Control PCA and see if the
error is resolved.
Refer to Section 5-11.