8.5.8 Data Structure on Memory Card
Fig. 8-46
Data structure on memory card
Flash
This is the folder where the backup file is being saved (to execute select I/0 –
Memory Card – Save backup).
The name of the saved file is always >Q_H_V_T.TXT<. It contains the internal
memory values on level, velocity, flow and temperature.
The file >DIAG.TXT< contains all messages including error messages which
might have been occurred during measurement operation. These might be start
and end of Internet communication, modem restart, CPU restart after system
reset or after reprogramming.
The respective message is labelled with date and time:
>:
received error/message
<:
reason of error/message cleared
The DIAG file is going to be saved only if you proceed as follows: I/0 – Memory
Card – save backup.
PARA
This folder includes all parameter files with a date stamp.
The content of this folder allows to retrace transmitter settings regarding the
measurement place as well as parameter settings which might have been modi-
fied.
The latest modification within the course of a day will be saved.
The file name is: PA TT MM JJ .TXT
(TT = day, MM = month; JJ = year)
NIVIDENT
The name of the measurement place.
If the name of the measurement place saved on card does not comply with the
name of the measurement place saved in the NivuSonic CO, the unit will
prompt to format the card. The NivuSonic CO will create a new file for meas-
urement values with the newly entered name as long as the card is not going to
be formatted.
Name of measurement
place.TXT
This is the file where the measurement values are saved. It is going to be saved
using the name of the measurement place set.
PARAMET.NIV
PARAMET.TXT
These files are created as soon as parameters are being saved on the memory
card. The file PARAMET.NIV is required in order to upload data to NivuSonic
CO. PARAMET.TXT is the print version of PARAMET.NIV as text file.
Use memory cards purchased from NIVUS only. Other manufacturer’s cards
may lead to irreversible loss of data or measurement failure (e.g. permanent
transmitter reset).